3–1
Rapid fFN 10Q
System User Manual
Section 3 — General Operating/Testing Instructions
HOLOGIC
Rapid fFN 10Q
SYSTEM DIAGNOSTICS
IN PROCESS
HOLOGIC
Rapid fFN 10Q
VERSION 2.0.0
03:00 PM 10/26/13
SECTION 3 — GENERAL OPERATING / TESTING INSTRUCTIONS
After instrument installation, the Rapid fFN 10Q Analyzer can be operated on a day-to-day basis by using the
following procedures. Read Section 4 for detailed descriptions of displays, prompts and operating sequences.
STARTING THE SYSTEM
1. Turn the analyzer power switch to the ON position. The power switch is
located on the left side of the instrument. (If the analyzer does not turn on,
see Section 7, Troubleshooting, Item 1.)
Upon power up, the analyzer displays the message
INITIALIZING
for 17 seconds. Then it beeps and the analyzer will perform
SYSTEM
DIAGNOSTICS
(a self-test of the analyzer components).
If there is a problem after the self-test, a beep will sound to indicate an error and an error code will be
displayed. If an error code is displayed, refer to the troubleshooting section of the manual. Verify the date and
time are correct. See Section 4, Software Functions — Detailed Descriptions, for setting Date/Time.
2. Once it is plugged in, the printer is continuously powered on. Ensure labels are in the printer. See
Section 6, Printer, for loading printer labels.