Section 7 — Troubleshooting
ITEM
PROBLEM
SOURCE
SOLUTION
10
Rapid fFN 10Q
QCette failed
during daily
quality control.
Rapid fFN 10Q
QCette or
Analyzer
Turn analyzer off and back on to reinitialize the system.
Confi rm that the QCette serial number on the device
matches the serial number programmed in the analyzer,
indicated on the Daily QC printout.
Ensure QCette is clean and not damaged. The QCette may
be cleaned with canned air.
Run 3-Daily QC. If this passes, no further troubleshooting is
required.
If Daily QC fails, perform 6-3-QCette setup. If QCette setup
is complete, run 3-Daily QC, followed by 8-Liquid Controls. If
both pass, no further troubleshooting is required.
If Daily QC fails a second time, call Technical
Support.