4–15
Rapid fFN 10Q
System User Manual
Section 4 —
Software Functions — DAILY QC
QCette Data Record — This record will be displayed on three screens. Each screen can be accessed by using
the ← and → keys. The complete record will be printed automatically if AUTOPRINT is set to ON, or it may be
printed/reprinted by pressing the PRINT/ENTER key.
QCette results are SYSTEM PASS or SYSTEM FAIL. The analyzer displays the PASS or FAIL result and the
fFN concentration for each QCette level (LVL1 and LVL2).
A FAIL result should be repeated. (See Section 7, Troubleshooting, Items 9 and 10.)
QCette
03:00 PM 10/26/13
SYSTEM: PASS
USER: JOHN SMITH
LVL1: PASS 20 ng/mL
LVL2: PASS 179 ng/mL
QCette SN:006460
CAL CODE:95HYA-76YTT
ANALYZER ID:G0237
ESC-MAIN MENU
Display and Printout
HOLOGIC Rapid fFN®
Rapid fFN 10Q QCette
SYSTEM: PASS
TIME: 03:00 PM DATE: 10/26/13
QCette LEVEL 1: PASS [ 20 ng/mL]
QCette LEVEL 2: PASS [179 ng/mL]
USER: JOHN SMITH
QCette SERIAL#: 006460
CALIBRATION CODE: 95HYA-76YTT
ANALYZER ID: G0237