5–1
Rapid fFN 10Q
System User Manual
Section 5 — Care of the Anaylzer
SECTION 5 — CARE OF THE ANALYZER
GENERAL CLEANING
Keep the analyzer free of dust. If needed, clean the exterior with a damp cloth and mild detergent.
WARNING:
Liquids MUST NOT be allowed to seep into the analyzer. Keep the analyzer dry at all times.
Liquids leaking into the analyzer may cause damage to the electrical components or possibly electrical
shock to the user.
CAUTION:
DO NOT use solvents of any type on any part of the analyzer. Solvents can damage the display and keypad.
CLEANING OF CASSETTE INSERTION SITE
The cassette insertion site can come into contact with biological fl uids and should be cleaned regularly.
CAUTION:
Use appropriate laboratory procedures for handling biohazardous materials.
CLEANING AGENTS APPROVED FOR USE
Reagents not listed below may cause discoloration to the analyzer case and membrane keypad.
The following cleaning agents may be applied with a cloth or lab wiper only. NEVER apply
agents by spray.
• 10% Bleach in water
• 75% Isopropyl Alcohol
• General laboratory cleaning detergent with disinfectant properties