4–17
Rapid fFN 10Q
System User Manual
Section 4 —
Software Functions — LIQUID CONTROLS
The analyzer automatically compares the CASSETTE LOT# used to set
calibration with the CASSETTE LOT # used for testing controls. If the
lot numbers do not match, the test process cannot continue. When this
occurs, the cassette lot number used for calibration will be displayed,
and the user is prompted to recalibrate the system.
From the CONTROL TEST MENU, select 1-LEVEL 1 control or
2-LEVEL 2 control.
Enter the control lot number and press ENTER. This fi eld will accept up
to 12 alphanumeric characters.
This message will be displayed if a cassette is present in the analyzer
prior to reaching the next screen. Remove cassette and press ENTER.
CASSETTE LOT CHANGED
CALIBRATE SYSTEM
L1002
ESC-MAIN MENU
CONTROL TEST MENU
1-LEVEL 1
2-LEVEL 2
LEVEL 1 CONTROL LOT #
>M1023
ENTER-CONFIRM
REMOVE CASSETTE
PRESS ENTER
TO CONTINUE