4–6
Rapid fFN 10Q
System User Manual
Section 4 —
Software Functions — TEST PATIENT
TEST PATIENT
Option 1 on the Main Menu screen allows the user to test a patient
sample. Follow the analyzer prompts.
The USER ID must be entered to proceed to the next step. Press ENTER
to accept the ID. This fi eld will accept 15 alpha or numeric characters.
The CASSETTE LOT# must be entered to proceed to the next step. For
convenience, the last 2 numbers only can be entered if the lot has not
changed. The CASSETTE LOT# is located on the cassette pouch. The
software requires that the lot number is entered in the correct format: one
alpha character followed by four numeric characters (e.g., L1002).
Note
: If the CASSETTE LOT# has changed, SET CALIBRATION must
be performed.
TEST PATIENT
USER ID
>JOHN SMITH
ENTER-CONFIRM
TEST PATIENT
CASSETTE LOT#
>L1002
ENTER-ACCEPT
fFN MAIN MENU
1-TEST PATIENT
2-SET CALIBRATION
3-DAILY QC