7–5
Rapid fFN 10Q
System User Manual
Section 7 — Troubleshooting
ITEM
PROBLEM
SOURCE
SOLUTION
9
Rapid fFN
10Q QCette
failed to
complete setup.
Rapid fFN
10Q QCette or
Analyzer
Turn analyzer off and back on to reinitialize the system.
Ensure QCette is clean and not damaged. The QCette can
be cleaned with canned air.
Repeat QCette setup as described in Section 4. If QCette
setup is complete, run 3-Daily QC, followed by 8-Liquid
Controls.
Do not bump or jar analyzer or QCette during the test.
If the QCette setup fails a second time, call Technical
Support.