7–11
Rapid fFN 10Q
System User Manual
Section 7 — Troubleshooting
ERROR CODE
DEFINITION
TROUBLESHOOTING PROCEDURE
1436
Low Input Voltage Detected
The correct power adapter is required. Ensure the 10Q power
adapter is being used with the analyzer. Contact Technical Sup-
port.
OD24
Possible QCette Problem
Turn analyzer off and back on to reinitialize the system.
Clean the QCette according to its directional insert.
Repeat QCette setup (refer to Troubleshooting Item 9).
OE21
High Light Level
Turn analyzer off and back on to reinitialize the system.
Identify if there is any new source of bright light (recent relocation
near a window, lab lights replaced, window blinds open). Remove
the source of excess light if possible or relocate the analyzer.
Refer to Troubleshooting Items 12 and 13. Retest Daily QC.
If Daily QC passes, continue with normal operation. If Daily QC
fails, contact Technical Support.
OE22, OE23
Possible Optics Problem
Turn the analyzer off and back on to reinitialize the system.
Retest Daily QC.
If Daily QC passes, continue with normal operation. If Daily QC
fails, contact Technical Support.
Other Codes
Call Technical Support.
If analyzer errors persist or recur, call Technical Support.