23
Programming and pairing of the foot control
Read this instruction before programming the foot control.
1.
Connect the foot control to the console using the cable. The Dental Delivery
System must be turned ON.
2.
Press and hold both ears down simultaneously until you hear three beeps
(about 5 sec.).
3.
Within 5 seconds activate one of the following:
a.
Choose with spring and linear pedal arm function: Activate the RIGHT
EAR shortly and you hear one beep.
b.
Choose with spring and logarithmic pedal arm function: Activate the
LEFT EAR shortly and you hear two beeps.
c.
Choose without spring and linear pedal arm function: Activate the LEFT
BRACKET shortly and you hear three beeps.
d.
Choose without spring and logarithmic pedal arm function: Activate the
RIGHT BRACKET shortly and you hear four beeps.
The encryption code is now transferred to the Receiver PCB.
4.
Within 5 seconds:
a.
Place the pedal arm at min. (to the left).
b.
Activate LEFT bracket shortly to select minimum position and you hear
one beep.
c.
Place the pedal arm at max. (to the far right).
d.
Activate RIGHT bracket shortly to select maximum position and you hear
5 beeps. The foot control is now paired with the unit and goes back to
normal work mode
If you do not press within 5 seconds or press something illegal, you will hear 10 rapid
beeps and foot control goes back to normal operating mode without having changed
anything
Right ear
Left ear
Right bracket
Left bracket
Pedal
arm
Min - Max
Battery indicator
Summary of Contents for Unicline S
Page 12: ...12 1 9 UnicLine S Labelled Diagram...
Page 54: ...54 12 5 Hydraulic diagram...
Page 64: ...64 13 1 Menu overview...
Page 83: ...83 Notes...