74
16.1
Probability of Occurrence & duration of
contact for parts.
Part
Time
Comment.
Instrument table
T < 1 s
Table 23, Is touched by the
operator for moving the table in
position, max 74°C
Turbine
10 s =< t < 1 min
Table 23, Handpice handled
during treatment by the operator,
max 51°C
Micromotor
10 s =< t < 1 min
Table 23, Handpice handled
during treatment by the operator,
max 51°C
Syringe
1 s =< t < 10 s
Table 23, Handpice handled
during treatment by the operator,
max 56°C
Lamp
1 s =< t < 10 s
Table 23, Handpice handled
during treatment by the operator,
max 56°C
Operating Lamp
T < 1 s
Table 23, Is touched by the
operator for moving the lamp in
position, , max 74°C
Suction tubes
10 s =< t < 1 min
Table 23, Handpice handled
during treatment by the operator,
max 60°C
Maximum Temperature
t < 1min.
Table 24, Applied part max 60°C