Universal variable foot control
The UnicLine S can be delivered with a universal variable foot control which is used
to control the selected instrument.
If your UnicLine S is supplied with a universal variable foot control, there will not be
3 pre-selected speeds. However, it can be programmed linearly or logarithmically -
please see the technical section for further details.
Setting the pedal spring.
The pedal spring can be activated or deactivated under the foot control by holding
the pedal arm in the MAX position while screwing the screw either out or in. (Only
turn the screw until it is free of the base plate).
If the spring option in the bottom is changed, remember to change the settings in
the program menu on the telescopic head.
Assistant call bracket
Left ear:
Brief activation of
the control (less
than 1 second)
changes the spray
selection for the
selected instrument.
Prolonged activation
(more than 1
second) activates
the mechanical chip
blow. Chip blow
remains activated
for the activated
time.
Assistant call bracket
Right ear:
Brief activation of the
control (less than 1 second)
changes the spray selection
for the selected instrument.
Prolonged activation (more
than 1 second) activates the
mechanical chip blow. Chip
blow remains activated for
the activated time.
Please see the
operating
instructions for the
UnicLine S patient
chair for a
description of the
joystick that
operates the chair.
The intensity of the selected instrument can be
increased or decreased by moving the pedal arm in
the direction of the arrow. To change the direction of
rotation of the motor, briefly press the pedal arm
down and then press the spray selector.
Min - Max
Summary of Contents for Unicline S