Turbine
If the equipment does not work as described below,
contact an authorized Heka Dental dealer immediately.
The following descriptions apply to the UnicLine S standard foot control. For
instructions on the UnicLine S variable foot control please see the instructions
specific to this.
Activation of the turbine
When the turbine is lifted from the instrument holder, the fibre optic is automatically
turned on. A constant press on the activation ring of the foot control starts the
turbine.
Modification of the spray function
In order to change the spray combination of the turbine, the instrument must be
lifted from the instrument holder. The spray combination is shown on the display. It
is indicated by spray symbol and the combination can be changed by briefly pressing
Indicates
spray air
ON
blue
Indicates
spray water
ON
green
If air/water spray is chosen, there will be an automatic chip blow for 2 seconds (with
the standard program).
For a powerful continuous mechanical chip blow, press and hold the middle ring. The
mechanical chip blow will continue as long as the press is maintained.
Adjustment of water and air
Regulators are installed under the turbine module:
Regulator
Regulation of:
Clockwise
Anticlockwise
Blue
Spray air
Decrease
Increase
Green
Spray water
Decrease
Increase
Neutral
Operating air Decrease
Increase
The driving air is adjusted during installation of the Dental Delivery System and
should not be readjusted.
There is a maintenance-free return air evacuation box in the console.
For more detailed information on the use of the turbine, please refer to the
attached instructions from the manufacturer.