50
12.2
Classification of Equipment
Please refer to the OEM manufacturer instructions for more
information.
Dental Delivery System
: Class IIA
all type B patient parts comply with
IEC/EN60601-1.
Method of operation
: Continuous operation with intermittent
load.
Protection against water
: IPX 0.
OEM Equipment Classification
(see OEM manufacturer instructions for more information)
UnicLine Chair:
Class I, type B.
Considered as applied part.
Multifunctional syringe
: type B.
Intermittent use
: 10 sec. ON / 20 sec. OFF.
Protection against water
: IPX 0.
Light polymerization lamp
: type BF.
Intermittent use
: 20 sec. ON / 40 sec. OFF.
Protection against water
: IPX 0.
Ultrasonic scaler
: type B
Operation with water
: 6 sec. ON / 3 sec. OFF for a maximum of 4 min.
with normal water supply from the unit. (1-5 bar)
Operation without water
: Intermittent use; 2 sec. ON / 18 sec. OFF for a
maximum of 10 min.
Protection against water
: IPX 0.
Micromotor
: type S3
Function in compliance with
: 3A for 60 sec. / 5A for 10 sec. Must be
CEI 34-1 type S
cooled down for 3 min. in the event of operation with
air or for 20 min. in the event of operation without
air.
Air cooling flowrate
: 10 l/min.
Protection against water
: IPX 0.
Operating
lamp:
Class I
Continuous operation:
No requirements for operating/inactive time.
Protection against water:
IPX 0.
Foot
control:
Class 1, type B.
Continuous operation:
No requirements for operating/inactive time.
Protection against water:
IPX 1.