38
9.0
Metasys automatic Separator
If the equipment does not work as described below,
contact an authorized Heka Dental dealer immediately.
To ensure the correct function and to avoid malfunction of the Metasys automatic
separator the following must be carried out:
Daily preventive maintenance
Rinse the suction hoses and the spittoon after every treatment. We recommend that
the spittoon and the suction system are cleaned every day at midday and in the
evening or refer to the manufacturer’s instructions for preventive maintenance to
keep the Dental Delivery System functioning properly.
We recommend the use of Metasys Green & Clean for cleaning (F-979).
Replacement of the collection tank
1.
Switch off the unit.
2.
Remove the cover plate furthest away from the patient chair.
3.
Lift up the locking bracket and pull out the entire tank element horizontally.
4.
Open the four clip locks. Lift the separator element up and down a couple of
times. Afterwards, clean the outside of the pump as well as the pump filter at
the bottom. Then, clean the three sensor electrodes (three black "sticks").
5.
Place the separator element in the new tank, facing the “front” label on the
tank.
6.
Close all four clip locks.
7.
Slide the tank element carefully along its rail and check that it is positioned
correctly. Lower the locking bracket.
8.
Switch on the Dental Delivery System and check that the green light-emitting
diode on the Separator keyboard indicator switches on.