19
Pairing of the foot control
Read this instruction before pairing the foot control.
1.
Connect the foot control to the console using the cable. The Dental Delivery
System must be turned ON.
2.
Press and hold the TOP and the middle ring down simultaneously until you
hear three beeps
3.
Within 5 seconds:
a.
Activate the TOP shortly. You hear one beep.
The encryption code is now transferred from the foot control to the receiver.
b.
Within 5 seconds:
c.
Activate the TOP shortly. You hear one beep.
d.
Activate the TOP again. You hear five beeps. The foot control is now
paired with the unit and goes back to normal work mode
If you do not press within 5 seconds or press something other than expected, you will
hear 10 rapid beeps and foot control goes back to normal operating mode without
having changed anything.
The Top, Middle ring and Activation ring can be activated from all sides
Increase
Decrease
Activation ring
Middle ring
Changes the spray /
Mechanical chip blow
Top:
OP-lamp ON/OFF
Battery indicator