59
Start deg. Stop
Must be set to 0 0 in UnicLine S
Setup Chair.
Press
F1
repeatedly to change between the options
Auto/None
.
Auto:
The chair runs automatically in the
rinse position upon activation of Rinse Position. In the Rinse Position you have these
functions on instrument table and joystick:
up (last position), left (stop).
None
: The chair does not run upon activation of Rinse position
Choices in this menu must be confirmed by pressing
OK
for the change to take effect.
Setup Foot ctrl.
Var! / Var
→
/ Pulse / Switch
Repeatedly press
F1
to change between the options Var! / Var
→
/ Pulse / Switch.
Lin/Log:
Press
F4
repeatedly to change between the options Lin / Log.
Name
Lin
(Linear control)
Log
(Logarithmic control)
Name
Var!
Variable foot controller without pullback spring.
Activate instrument by pressing the pedal arm.
Var→
Variable foot controller with pullback spring .
Activate instrument by horizontal pressure on the pedal arm.
Pulse
Round variable foot controller (not wireless).
Switch
Standard foot controller and standard wireless foot
controller.