WARNING:
For use by qualified and trained personnel only.
WARNING:
Do not install where there is a risk of explosion. The UnicLine S Dental
System is not intended for operation in oxygen rich environments or in the presence
of flammable anaesthetics or gases.
WARNING:
Clean, disinfect, and sterilize new or repaired handpieces and
instruments before first use and between each patient use. Only use sterilized
handpieces and instruments during treatment. Non-sterilized handpieces and
instruments may cause bacterial or viral infections. Always sterilize handpieces and
instruments after operation.
Please see the section on approved cleaning agents and methods
for a detailed description of cleaning methods and maintenance for the
UnicLine S. See enclosed OEM instructions for cleaning and maintenance of
any OEM equipment and instruments.
WARNING:
Always operate high-speed handpieces with water coolant. Operating a
high-speed handpiece without water coolant can cause thermal injury to the patient.
WARNING:
Do not use this device for dental implant procedures.
WARNING:
The water used by the Unic unit's instruments and cup-filler is only
intended for rinsing.
WARNING:
The mains tap must be switched off when the Dental Delivery System is
not in use.
Electromagnetic Compatibility (EMC)
Changes or modifications to this product not expressly approved by Heka Dental A/S
may result in increased emissions or decreased immunity performance of the product
and could cause EMC issues with other equipment. This product is designed and
tested to comply with applicable regulation regarding EMC and shall be installed and
put into service according to the EMC information stated below:
WARNING:
Use of portable phones or other portable or mobile radio frequency (RF)
emitting equipment near the product may cause unexpected or adverse operation.
WARNING:
The use of accessories, transducers and cables other than those supplied
may result in increased emissions or decreased immunity performance of the
product. The product shall not be used adjacent to, or stacked with, other equipment.
If adjacent or stacked use is necessary, the user shall be responsible to test to verify
normal operation in the configuration in which it is being used. Detailed information
on electromagnetic interference related to the UnicLine S begins on page 71 of this
manual. See page 74 for recommendations for distances to be maintained between
the UnicLine S and other electronic devices.