21
Programming and pairing of the foot control
Read this instruction before programming the foot control.
1.
Connect the foot control to the console using the cable. The Dental Delivery
System must be turned ON.
2.
Press and hold the TOP and the middle ring down simultaneously until you
hear three beeps.
3.
Within 5 seconds activate one of the following:
a.
Choose linear pedal arm function: Activate the TOP until you hear two
beeps.
b.
Choose logarithmic pedal arm function: Activate the MIDDLERING until
you hear one beep.
The encryption code is now transferred from the foot control to the receiver.
c.
Within 5 seconds:
d.
Without activating the pedal arm (arm to the left)
e.
Activate the TOP shortly. You hear one beep.
f.
Move the pedal arm to max. at a steady speed (hold arm to the right)
g.
Activate the TOP again. You hear five beeps. The foot control is now
paired with the unit and goes back to normal work mode.
If you do not press within 5 seconds or press something illegal, you will hear ten
rapid beeps and foot control goes back to normal operating mode without having
changed anything.