49
12.1
Installation Requirements
For positioning of the UnicLine S, please refer to the Installation instructions,
Section 6.
1.
230 Volt/115 Volt +/- 10 %, 50/60 Hz, earthed. Branch fuse, 10A/20A.
2.
Water, inlet pressure 2-6 bar (30-90 psi) Ball valve with R 3/8" inner
thread Flow rate <- 4 l/min. (Max. consumption for the whole area.)
Water quality: <- 8 dH (1 dH = 20 mg Ca/3l water)
Top 5 cm above the floor.
3.
Air, inlet pressure 4.5 - 6 bar (65-88 psi). Ball valve with R 3/8" inner
thread Flow rate <- 55 l/min. (Max. consumption for the whole area.)
Humidity limit > 20 ºC at atp.
Oil free (max. 0,5 mg/m³)
Particulate contamination limit not > 100 ppm/m³ for 1µm to 5µm particle
size.
Top 5 cm above the floor.
4.
Outlet, 32 mm exterior diameter, PVC tube. Top 5 cm above the floor.
Capacity: min. 10 l/min
5.
Suction, 32 mm exterior diameter, PVC tube. Top 5 cm above the floor.
(DN40)
Vacuum: >- 150 mbar, Flow rate >- 550 l/min.
6.
Control cable for suction motor, 2 x 1.5 mm
2
. Free end 50 cm above the
floor.
7.
Control cable for assistant call, 2 x 1.5 mm
2
. Free end 50 cm above the
floor.
8.
Cable for x-ray equipment, if applicable. Cable type/number depends on
brand. Free end 50 cm (at least) above the floor.
9.
Temperature and humidity:
During operation: +10 ºC to +35 ºC, non-condensing atmospheric
humidity from 20 – 75 %, pressure 800 hPa – 1060 hPa.
Storage/transport: 10 ºC to +40 ºC, non-condensing atmospheric
humidity from 50 – 80 %, pressure 500 hPa – 1060 hPa.
230V/115V output to chair, light and extra equipment. All are connected in parallel
and the total load must not exceed 5A/1150 W / 10A/2300 W. All of these outlets
have a T5.0A/T10.0A fuse.
Warning!
In order to prevent electric shock, the UnicLine S must only be connected
to a mains supply with protective earth
.