P/N 53063-50, Manual Revision: B
Sequestering on the Cell Saver 5+ System
7-17
6a. Observe the centrifuge bowl filling and allow air from the circuit and the
bowl to flow to the air bag or PPP collection bag.
b. When plasma reaches the effluent line, the machine will decide if the
process should pause as determined by the selection that was made for air
management.
If the selection to manage the system air was in the AIR bag, the process
stops the pump and prompts the operator to open the YELLOW (PPP) bag
clamp and close the air bag clamp. Once this is completed the machine
waits for the START key to be pressed to continue processing as shown in
the following screen display:
Caution:
Never close the yellow (PPP), blue (PRP), and white (air) bag clamps at
the same time.
If the PPP collection bag was chosen to manage the system air, the machine
continues processing.
PAUSE
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press START to resume collection of PPP
SEQUESTER
139 ml
0 ml
0
Pump Rate:
60
Centrifuge:
4750
Press MODIFY to change parameters
ml/min
rpm
PPP Volume:
PRP Volume:
0 ml
0 ml
0/
0/
OPEN YELLOW bag clamp
CLOSE AIR bag clamp