P/N 53063-50, Manual Revision: B
Sequestering on the Cell Saver 5+ System
7-21
has been reached the pump restarts and the process repeats. This process of
stopping and starting will continue until the PRP collection is completed.
11. After 10 ml of PRP have been collected the effluent line sensor starts looking
for RBCs spilling into the PRP collection bag. Once RBCs have been
detected the machine alerts the operator that the PRP Extended Collection
has started. The PRP Extended Collection Volume is programmed by the
operator in the Sequester Setup.
Note: If the sensor tubing is not properly installed in the line sensor, the CS5+ will
not detect the red blood cells spilling into the PRP collection bag and the CS5+
will not stop at the desired end collection point.
If the line sensor does not sense the red blood cells spilling into the PRP
collection bag the operator can press the stop or empty key to end the PRP
collection.
125
ML
PRP
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press SELECT to advance highlight
SEQUESTER
269 ml
0 ml
0
Pump Rate:
60
Centrifuge:
4750
Press ARROW keys to change values
ml/min
rpm
PPP Volume:
PRP Volume:
100 ml
30 ml
100/
30/
Press MODIFY to change parameters
Extending PRP Collection
COLLECTION