Sequestering on the Cell Saver 5+ System
P/N 53063-50, Manual Revision: B
7-10
8. When the bowl completely empties, the pump will stop and the following
display will be shown:
9. The RBCs are reinfused to the patient as deemed necessary by anethesia.
Refer to AABB standards for expiration time of RBCs.
Warning: In accordance with FDA guidelines and AABB standards a transfusion
filter capable of retaining particles potentially harmful to the recipient should
be used when returning processed concentrated red cells.
10. If another pass is desired, the entire procedure may be repeated using Steps
1 through 7, until the desired volume of PPP/PRP has been collected.
Note: The concentrate mode can be used if it is necessary to pull additional blood
into the bowl for the final cycle. Just press the concentrate key if you run out of
whole blood when filling the bowl. This allows the operator to push out the
remaining PRP from the bowl.
STANDBY
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press START to fill bowl or CONC to concentrate
SEQUESTER
510 ml
125 ml
1
Press MODIFY to change parameters
Press YES to select Cell Saver protocol