Sequestering on the Cell Saver 5+ System
P/N 53063-50, Manual Revision: B
7-20
9. Once the START key is pressed the machine first looks at the centrifuge
speed for PRP collection that was programmed in Sequester Setup. If the
programmed PRP collection centrifuge speed is equal to or greater than the
programmed Fill Bowl centrifuge speed, the pump is started and PRP
collection is resumed. If the programmed PRP collection centrifuge speed is
less than the programmed Fill Bowl centrifuge speed, there will be a 30
second delay to allow the centrifuge to reduce its speed to reach its new
programmed speed at which time the pump will restart and PRP collection
will resume as shown in the following display:
10. Once the PRP collection process resumes, PRP is being collected into the
PRP collection bag and the PRP Volume on the display is incrementing
along with the Processed Volume.
If the “Milk PRP Collection Layer?” parameter in the Sequester Setup is set to YES,
the pump will process the “Milk Vol” parameter setting and then the pump will
stop and pause for the “Milk Delay” parameter setting. Once the Milk Delay time
125
ML
PRP
Reinfusion Vol:
Bowls Processed:
Press SELECT to advance highlight
SEQUESTER
0 ml
0
Pump Rate:
0
Centrifuge:
2400
Press ARROW keys to change values
ml/min
rpm
PPP Volume:
PRP Volume:
100 ml
0 ml
100/
0/
PUMP RESTART IN XX SEC
Press START to override
Press MODIFY to change parameters
COLLECTION
125
ML
PRP
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press SELECT to advance highlight
SEQUESTER
248 ml
0 ml
0
Pump Rate:
60
Centrifuge:
4750
Press ARROW keys to change values
ml/min
rpm
PPP Volume:
PRP Volume:
100 ml
9 ml
100/
9/
Press MODIFY to change parameters
COLLECTION