P/N 53063-50, Manual Revision: B
Appendices
APP-13
IEC 60601-1-2: 2001, Table 204: Guidance and manufacturer’s declaration - electromagnetic immunity
The Cell Saver system is intended for use in the electromagnetic environment specified below. The customer
or the user of the Cell Saver system should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment guidance
Portable and mobile RF communication
equipment should be used no closer to any
part of the
Cell Saver
system, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
(To be calculated)
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
V1=3 Vrms
E1=3 V/m
Where
P
is the maximum output rating of
the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation distance in meters.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey
a
should be less than the compliance
level in each frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio AM and FM broadcast and TV broadcast cannot be predicted theoretically with accuracy. To asses the electromag-
netic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength
in the location in which the Cell Saver system is used exceeds the applicable RF compliance level above, the Cell Saver system
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as reorienting or relocating the Cell Saver system.
b)
Over the frequency range 150 kHz to (0 MHz, field strengths should be less than
d
3 5
,
V
1
---------
P
=
d
3 5
,
E
1
---------
P
=
80 MHz to 800 MHz
d
7
E
1
------
P
=
800 MHz to 2,5 GHz
V
1
[ ]
V/m.