Manual Operation
P/N 53063-50, Manual Revision: B
6-6
During the WASH mode the following display will be shown:
Note: If during the WASH mode the red cell/supernatant interface disappears and
red cells begin to spill into the waste bag, reduce the pump speed to zero or press
pump pause and allow this interface to reappear. Slowly increase the pump speed
to a rate that just maintains this interface (usually 200 ml/minute wash or greater).
6. Press the EMPTY key (the recommended maximum EMPTY speed is 500 ml/
min for the 225 and 125 ml bowls and 100 ml/min for the 70ml bowl).
During EMPTY the display will read:
A “Centrifuge Stopping” message will blink on the screen until the centrifuge
comes to a complete stop. The concentrated red cells are pumped from the bowl
to the reinfusion bag. Under certain conditions the centrifuge will briefly rotate
in both directions to expedite the removal of the red cells from the bowl.
225
ML
WASH CELLS
Processed Vol:
Wash Vol:
Reinfusion Vol:
Bowls Processed:
Wash Volume: 150 ml
Press MODE key for AUTOMATIC
MANUAL
800 ml
150 ml
0 ml
0
Press ARROW keys to change values
NaCl
Pump Rate: 500 ml/min
225
ML
EMPTY BOWL
Processed Vol:
Wash Vol:
Reinfusion Vol:
Bowls Processed:
Press MODE key for AUTOMATIC
MANUAL
1250 ml
1000 ml
75 ml
0
Pump Rate: 500 ml/min
RBC
Press ARROW keys to change values