Appendices
P/N 53063-50, Manual Revision: B
APP-8
A
PPENDIX
C: M
ACHINE
MESSAGES
/
TROUBLESHOOTING
Machine Messages/Troubleshooting
Condition/Reason
Mode
Action
“AIR DETECTED EARLY”
Air detected before bowl has been
emptied.
EMPTY/RETURN
1. Check tubing placement.
2. Check for occlusions in effluent tubing.
3. Check for bowl displacement air in waste
bag.
“AIR DETECTED IN FILL CYCLE”
FILL
1. Check tubing in air detector.
2. Check for occlusions in harness and
bowl.
3. Open clamp on stepdown connector to
red line.
4. Check suction: less than 200 mmHg.
5. Check tubing in valves for correct place-
ment.
“BOWL TYPE CONFIRMATION”
The white, 70 ml bowl adapter has
been detected in the centrifuge
well.
FILL/LOAD
DISPOSABLE
1. Press YES if using the white bowl adapter.
2. Press NO if NOT using the bowl adapter.
“CENTRIFUGE COVER NOT
CLOSED”
Centrifuge does not operate unless
cover is closed.
FILL/WASH
1. Open and close centrifuge cover.
2. Ensure inlet and outlet tubing is placed
correctly.
“CENTRIFUGE ERROR”
Overspeed detected.
Unable to reach or maintain rate.
Failed to stop.
CENTRIFUGE STALLED
FILL/WASH
Check bowl for proper seating.
“CENTRIFUGE STOPPING”
Delay while centrifuge gradually
slows to a stop.
STOP/EMPTY
System ignores any input from the control
panel.
“CENTRIFUGE WARNING”
Fluid detected in centrifuge well.
ANY
1. Inspect disposable inside centrifuge well
for leakage.
2. Check effluent tubing for occlusion.
3. Check if waste bag is too full to permit
more fluid.
4. Clean centrifuge well as necessary.