Sequestering on the Cell Saver 5+ System
P/N 53063-50, Manual Revision: B
7-24
15. When the bowl completely empties, the pump stops and the following
display is shown:
16.
The RBCs should be treated the same as a unit of washed, concentrated
RBCs in terms of administration to the patient and outdate of the product.
Warning: In accordance with FDA guidelines and AABB standards, a transfusion
filter capable of retaining particles potentially harmful to the recipient should
be used when returning processed concentrated red cells.
17. If another pass is desired, the entire procedure may be repeated using Steps
5 through 14, until the desired volume of PPP/PRP has been collected.
Note: The concentrate mode can be used if it is necessary to pull additional blood
into the bowl for the final cycle. Press the CONC key if you run out of whole
blood when filling the bowl. This allows the operator to push out the remaining
PRP from the bowl.
STANDBY
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press START to fill bowl or CONC to concentrate
SEQUESTER
299 ml
199 ml
1
Press MODIFY to change parameters
PPP Volume:
PRP Volume:
40 ml
60 ml
0/
0/
Press YES to select Cell Saver protocol