P/N 53063-50, Manual Revision: B
Appendices
APP-9
“CENTRIFUGE WARNING”
Fluid detected at base of bowl.
1. Power OFF.
2. Clean centrifuge.
3. Restart machine.
“CLAMPED LINE DETECTED”
Pressure sensed in fluid lines.
ANY
1. Check for occlusions or kinks in tubing.
2. Ensure clamps on reinfusion bag and res-
ervoir are open.
DISPLAY INOPERATIVE
ANY
Contact Haemonetics Hotline.
“LONG EMPTY (or RETURN)
CYCLE“
Air should be detected by 300ml.
EMPTY/RETURN
1. Check tubing placement.
2. Check for occlusions in effluent, blue and
red tubing.
3. Check for bowl displacement air in waste
bag.
4. Verify bowl size recognition.
“PLEASE BE SURE PUMP PLATEN
AND VALVE DOOR
ARE CLOSED”
ANY
Inspect closure of pump and valve door.
“PRODUCT BAG WAS EMPTIED”
Air is detected.
CONC
1. If bag is empty, press STOP to wait for
more salvaged blood or press WASH to
complete cycle.
2. If bag is not empty:
!
Check tubing in air detector.
!
Check tubing in valves for correct
placement.
“PUMP ERROR”
Pump is not turning.
Incorrect pump rate is detected.
ANY
1. Check pump platen.
!
Open pump platen.
!
Check for proper tubing placement.
!
Close pump platen.
2. Clean pump rotor. (Refer to Appendix B
in the Operator’s Manual.)
“PUMP RESTART IN XX
SECONDS”
Centrifuge regains rate before
starting pump to prevent RBC
spillage.
FILL/WASH
No action required.
May override after 15 seconds, IF RBCs have
reconcentrated, by pressing START.
“SALINE BAG EMPTY
Air is detected during the WASH
mode.
WASH
1. If bag is empty, add more wash solution.
2. If bag is not empty:
!
Check spike in bag.
!
Open clamp to bag.
!
Close clamp to unused spike.
!
Check tubing in air detector.
!
Check tubing in clamps for correct
placement.
Machine Messages/Troubleshooting
Condition/Reason
Mode
Action