Sequestering on the Cell Saver 5+ System
P/N 53063-50, Manual Revision: B
7-6
4. Press MODIFY to save changes and return to the current settings screen.
5. Press START to enter the STANDBY screen.
Note: Performing Sequestering is a MANUAL operation requiring operator
intervention to move from one stage to another. The process should be
understood completely before pursuing.
Note: Before starting the sequestering process, check to make sure that the AIR
bag clamp is open and the collection bag clamps are closed, unless using the PPP
bag for air management. Ensure that while the bowl is filling, the air in the
disposable system is entering the AIR bag. The air will be emptied out of the AIR
bag during the empty mode.
Note: If pump pause is selected during processing, the pump will stop and the
centrifuge speed will be maintained for five minutes. After five minutes, the speed
will be reduced and after six minutes in pause the centrifuge will stop.
STANDBY
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press START to FILL bowl or CONC to concentrate
Press MODIFY to change parameters
SEQUESTER
0 ml
0 ml
0
Press YES to select Cell Saver protocol