Automatic Operation: Cell Salvage
P/N 53063-50, Manual Revision: B
5-14
Once the MODIFY key is pressed a screen similar to the following is displayed:
1. To enable the Empty RBCs to Red Line option press the SELECT key until the
“Empty RBCs to” Option is highlighted and then press the up arrow to
indicate the Red Line.
At this point a screen similar to the following will appear:
2. It is necessary to designate where the volume of washed RBCs will be
pumped after the red line — to the reservoir or to the circuit. This
information will allow the CS5+ to accurately account for the volume
processed and reinfused to the patient.
3. Press MODIFY to return to the STANDBY screen.
CELL SAVER SETUP
Press ARROW keys to change values
Press SELECT to advance highlight
Min Wash Vol. - 225:
-125:
Reservoir Level:
Resume at Level:
Reservoir Level Sensor:
Pump Regulation:
Empty RBCs to:
AUTOMATIC
Press MODIFY when finished
NO WASH Option:
OFF
1000 ml
750 ml
800 ml
400 ml
ON
ON
Blue Line
CELL SAVER SETUP
Press ARROW keys to change values
Press SELECT to advance highlight
Min Wash Vol. - 225:
-125:
Reservoir Level:
Resume at Level:
Reservoir Level Sensor:
Pump Regulation:
Empty RBCs to:
AUTOMATIC
Press MODIFY when finished
NO WASH Option:
OFF
1000 ml
750 ml
800 ml
400 ml
ON
ON
Red Line
Volume to:
Reservoir