P/N 53063-50, Manual Revision: B
Setup Procedures for Cell Salvage
4-7
2. Ready the Machine and Bowl Disposable.
!
Open the fluid deck cover, pump lever, and manifold door. Open the
centrifuge cover by pulling the grey/blue knob on the right side of the
cover and opening the header arm.
!
Peel back the protective lid on the tub.
3. Install the Bowl
!
Lift the bowl out of the tub, center and place it firmly into the centrifuge
well or into the white centrifuge adapter ensuring the red indicator lines
are visible (70m ml bowl). Push the bowl (or adapter) down and snap into
place. Ensure that the lower bowl port points to the right of the machine.
Figure 4-8: Install Bowl
!
Swing the bowl arm clockwise into place, and secure the bowl header by
turning the locking knob from 8 o’clock to 12 o’clock.
!
Manually spin the bowl to ensure proper seating and level rotation.
Figure 4-9: Secure Header Arm