Equipment and Disposable Description
P/N 53063-50, Manual Revision: B
2-8
Automatic Controls (See Figure 2-4.)
Located directly below the display, the Automatic controls regulate the automatic
functioning of the CS5+. The Automatic Controls are as follows:
a. START, as its name suggests, is used to initiate the first cycle of a set of cycles.
Also, if the system is in STANDBY, pressing START resumes the process
where it was stopped.
b. STOP simply stops the pump and centrifuge and closes all valves. The
machine “remembers” which action it was performing when STOP was
pressed. Pressing STOP twice enters STANDBY. Pressing START will cause
the CS5+ to resume processing in automatic or manual.
c. MODE (Auto/Manual) Pressing this key once while in Automatic operation
will move the system into Manual control. The LED will illuminate next to
the word MANUAL or AUTO to indicate the mode that the system is in. Also,
the available function keys will be backlit and made available for the
operator to press. The operator can easily revert to Automatic control by
pressing the MODE key to switch back to AUTO processing.
Pressing this key twice within two seconds while in Automatic operation or
once when in Manual operation allows the operator to enter an emergency
option. A message requesting confirmation to enter the emergency option
will be displayed on the screen. If, within five seconds, the operator does not
press YES to confirm the emergency option, the CS5+ will revert to the
AUTOMATIC mode.
d. HELP assists the user with helpful information as previously discussed.
Manual Controls (See Figure 2-5)
Located below the Automatic Controls on the left, these keys allow the operator
to manually run the machine once MANUAL operation is selected by pressing
the MODE key. These keys may also be accessed during the AUTOMATIC mode.
Modify Controls (See Figure 2-5)
Located below the Automatic Controls on the far right, these keys allow the
operator to modify the set parameters for the given process or permanently. The
Modify keys become active during protocol selection, Protocol Setup, or various
points during both Automatic and Manual operation.