Setup Procedures for Cell Salvage
P/N 53063-50, Manual Revision: B
4-12
Note: Changes made at this point will become permanent until changed by the
user. The no wash option, pump regulation off option, and min wash volumes
below 1000 ml for a 225 ml bowl, 750 ml for a 125 ml bowl are retained on a
per procedure basis only.
The selection of the sequestering protocol should be made at this point. The
sequestering protocol is not available for use with the 70ml bowl.
If sequestering is chosen, the following screen will appear. If no changes are to
be made, press START. For further information about the sequestering protocol
refer to Chapter 7.
If the Cell Saver protocol is chosen, after all modifications have been completed,
press the MODIFY key to return to the current settings screen. Then press START
to proceed to STANDBY. The following screen will be displayed.
Press START to FILL bowl, CONC to concentrate, or wait for the reservoir level
sensor to sense approximately 800 ml in the reservoir.
CURRENT SETTINGS
Press START to begin procedure
Press MODIFY to change settings
MANUAL
Press YES to restore default settings
Assisted Sequester Mode
OFF
60
4750
NO
ml/min
rpm
Fill Bowl DRAW Rate:
Fill Bowl Centrifuge Speed:
Return Plasma?
STANDBY
Processed Vol:
Wash Vol:
Reinfusion Vol:
Bowls Processed:
to FILL automatically
Press START to FILL bowl or CONC to concentrate
Press MODIFY to change parameters
AUTOMATIC
0 ml
0 ml
0 ml
0
Need 800 ml in reservoir