P/N 53063-50, Manual Revision: B
Appendices
APP-7
A
PPENDIX
B: C
LEANING
PROCEDURES
Cleaning Procedures for Cell Saver 5+ Parts (detailed instructions in chapter 8)
Part
Procedure
Air Detector
1. Floss with lint-free cloth dampened with clear water.
2. Floss again using a dry lint-free cloth.
Air Filter
1. Remove the four screws on the louvered panel covering the air filter. Lay the
panel door and screws aside.
2. Pull out filter and rinse with warm water until clean. Do not use soap or cleaning
solutions.
3. Let the filter dry over night. Replace the filter and panel door.
Centrifuge Spills:
1. Wipe clean the centrifuge chuck and centrifuge well with a damp cloth using a
mild detergent spray.
2. Ensure mechanical locking fingers are thoroughly cleaned.
3. Dry all areas with a lint-free cloth.
4. Remove and replace the biohazard waste bag, if necessary.
Control Panel
1. Wipe the control panel with a mild detergent-dampened cloth.
2. Wipe with a lint-free cloth dampened with clear water,
3. Dry with a lint-free cloth.
Effluent Line
Sensor
1. Floss with a lint-free cloth dampened with clear water.
2. Floss again using a dry lint-free cloth.
Bowl Optics
Sensors
1. Clean daily and when soiled to ensure adequate detection of and Fluid Leak Sen-
sors reflective light.
2. Wipe the sensor with a lint-free cloth dampened with clear water.
3. Dry completely using a lint-free cloth.
Pump
1. Unscrew the pump cover and remove the pump rotor.
2. Wipe inside of the pump rotor assembly with a mild detergent-dampened cloth.
3. Rinse the pump under a faucet with warm water. If necessary, pump rotor can be
placed in an ultrasonic cleaner.
4. Dry thoroughly using a lint-free cloth and replace the pump rotor.
Tubing Clamps and
Clamped Line
Sensor
1. Wipe with a lint-free cloth dampened with mild detergent.
2. Wipe with a lint-free cloth dampened with clear water.
3. Dry with a lint-free cloth.