Chapter 5: Troubleshooting
Servo Controlled Oxygen
5-16
Service Manual
Item
Description
Oxygen0
Oxygen reading in percent from first sensor cell:
Oxygen1
Oxygen reading in percent from second sensor cell:
Sensor0
Voltage output in millivolts from first sensor cell.
Sensor1
Voltage output in millivolts from second sensor cell.
Note:
The following information applies to the oxygen0, oxygen1,
sensor0, and sensor1 values:
•
5 - 90 mv = 21 - 100% O2.
Temp
Temperature in degrees Celsius read from a sensor inside the sensor
housing located beneath the bed. Refer to temperature to resistance
curve in the Tips section (
Humidity
Relative humidity in percent read from a sensor inside the sensor
housing located beneath the bed.
FIGURE 5-1. Servo 02
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...