Chapter 7: Repair Procedures
Left Rail Internal Repairs
7-28
Service Manual
7.5.2 Replacing the Rail Buoyancy Springs
“Figure 7-29 Left Rail Springs and Spools”
.
1.
Using a 4 mm hex key, remove the screw that secures the springs to the lift rail. The springs will roll back
around their spools.
a.
The spools slide out of their mounting slots. To reinstall the spools, assemble each of the spools by
wedging them into the rail with the plastic washers at either side of each spool, then slide all the
assembled spools into their mounting slots at the same time. Lubricate the spring spools by
placing a drop of Lubriplate where the spool hub seats in the slot.
Note:
Replace all the buoyancy springs in an upright at the same.
STEP 6
FIGURE 7-29. Left Rail Springs and Spools
Buoyancy spring
Plastic washer
Spring spool
Cable carrier channel
Cable carrier
Lift rail
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...