Chapter 6: Alarms and System Failure Codes
System Failure Codes
Service Manual
6-15
6.2.8 System Failure 8
System failure 8 indicates: No Line Frequency. When troubleshooting this error code, have control board, 50
pin ribbon cable, and relay board on hand.
6.2.9 System Failure 11
System failure 11 indicates: The ADC on the control board is not operating to spec (too slow).
6.2.10 System Failure 12
System failure 12 indicates: Incorrect control board system software checksum.
Cause(s)
Action(s)
Defective control board
Measure the signal on the control board between J9 pin 43, and TP1-4
(Ground). It should be a 120 hertz signal. If the signal is OK, replace the
control board.
Defective 50-pin ribbon
cable between the relay
board and control board
If no signal, check continuity on the 50-pin cable, pin 43. If defective,
replace the cable.
Defective relay board
Replace the relay board.
Cause(s)
Action(s)
Defective humidifier
sensor or cable
Disconnect J1 on the control board. If it powers up OK, either the cable or
the humidity sensor is defective. Reconnect J1 and disconnect the 4 pin
humidity sensor connector in the compartment probe. If the device now
powers up OK, the humidity sensor is defective. If not, the cable is
defective.
Defective control board
If the error persists with the replaced sensor or cable, replace the control
board.
Cause(s)
Action(s)
Defective system software
PROM on control board
Install new system software 2089261-001.
Defective control board
If the error persists with new software, replace the control board.
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...