Chapter 3: Calibration and Maintenance
Scale Calibration
3-6
Service Manual
Note:
One 5 kg Certified Test Weight (part number 6600-0209-800) is required to perform the calibration
procedure.
1.
Clear all objects from the bed, and ensure only the mattress and clear plate remain on the scale.
2.
Ensure the bed is level.
3.
In user mode, touch
Setup
on the Home screen. The Setup screen appears.
4.
Touch the data Baby Name field
Enter here
. The keyboard appears.
5.
If this is the first time accessing the application, enter
Pass0000
and go to step 6. If not, enter your
password (pa PIN). The password is not case sensitive and may be typed in upper or lower
case.
6.
Wait for
Password Verified
to be displayed.
7.
Turn off the Standby switch to turn off the device.
8.
Turn on the Standby switch within one minute to turn on the device. The device restarts in service mode
and is open to the Setup screen.
9.
In service mode, touch
Calibration
in the left-navigation area. The display area is populated.
10.
Touch
Cal Scale
.
11.
When the screen prompts
Remove the Weight and Press Next
, verify the weight is removed and press
Next
.
A series of instructions appears on-screen.
FIGURE 3-3. Cal Scale
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...