Chapter 7: Repair Procedures
Removing the Upper End Cap Spring
7-44
Service Manual
7.8 Removing the Upper End Cap Spring
“Figure 7-44 Top Rail End Cap”
1.
To replace the retaining spring, use a punch to push the roll pin in the top of the lift rail out about ¼ inch
and remove the old spring.
2.
Remove the old spring from the end cap.
3.
Use a needle nose pliers to hook one end of the new spring onto the roll pin.
4.
Push the roll pin back into the rail with the new spring attached.
5.
Hook the other end of the new spring with a 1.5 or 2 mm “long” hex key and stretch it up into the shaft
of the end cap. Hook it on the roll pin inside the end cap shaft.
FIGURE 7-44. Top Rail End Cap
Spring
Top inner end cap
Top outer end cap
Roll pin
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...