Chapter 4: Understanding the Service Interface
Upgrade
Service Manual
4-7
The USB and Ethernet connectors are identified with the electrostatic discharge sensitive label on the
back of the display assembly. The connectors should not be touched, nor should any connector be
plugged in until all possible electrostatic charges have been discharged by either touching the rails or
other metal surface on the device.
2.
Insert the USB drive with the new software into the USB Port.
3.
Wait until Software Validation is complete.
4.
Select
Upgrade
or
Reinstall
.
Upgrade
updates software components that have changed.
Reinstall
updates all software
components.
5.
Follow-on-screen instructions.
Note:
Software upgrade will fail if the Alarm Board is not fully functional.
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...