Chapter 7: Repair Procedures
Display Module Repairs
Service Manual
7-69
6.
Working from the front of the device, using a T10 torx tool, remove the two (30 x 10) thread forming
screws that mount the right hand side of the LCD display assembly to the display stand offs.
.
7.
Slide the display to the right to remove it from the left hand standoffs. Gently set the display on top of
the temporary support (OmniBed) or Incubator hood to reduce the stress on the harnesses while
accessing the harness connections to the SBC PCA.
Caution:
Be sure to support the display during disassembly to avoid damaging the display.
Caution:
The leads on the Display Harnesses are fragile. Care when removing the Display
Assembly is warranted.
FIGURE 7-66. Remove Thread Forming Screws
FIGURE 7-67. Set Down Display
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...