Chapter 6: Alarms and System Failure Codes
Alarms
6-10
Service Manual
6.1.18 Motor Drive Failure Alarm
6.1.19 Power Failure (LED Indicator)
Cause or Circumstances
Failure
Action
The software has detected that
the elevating base motor or the
canopy motor was running
when not turned on.
Defective relay board
Replace the relay board.
Note:
Humidifier will not operate
during this failure.
Cause or Circumstances
Failure
Action
No AC power into device
On/Off switch is not turned on.
Verify cord is plugged into a live
wall outlet. Verify power switch/
circuit breaker on back of device is
turned on.
Mains power is present but the
5V power supply is not present.
Defective fuse F1 or F2 on relay
board
Check the 5 volt supply. If no 5 volts,
possible defective fuses on mains
to the power supply, check F1 and
F2 on relay board.
Defective power supply
Check input to power supply for
mains voltage. If present and no 5
volts at output, then power supply
is defective.
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...