Chapter 7: Repair Procedures
Display Module Repairs
7-72
Service Manual
4.
Working from the back of the device using a T10 torx tool, remove the eight (M3 x 8) thread forming
screws that mount the EMI shield to the carrier plate and carefully remove the EMI shield.
5.
Working from the back of the device firmly hold the SBC PCA then using a T10 torx tool remove the six
(M3 x 8) thread forming screws that mount the Display Driver to the carrier plate.
6.
Remove the SBC PCA.
FIGURE 7-72. Remove EMI Shield
FIGURE 7-73. Remove SBC