Chapter 4: Understanding the Service Interface
Setup
4-4
Service Manual
Note:
Resetting maximum preheat level to above 25% will result in noncompliance to device standard IEC
60601-2-21.
4.4.2 Pedals
The Bed/Canopy can be operated from the service screen when pedals are not working by touching the On/
Off toggle for each item. In the event of a bed raise/lower or canopy foot pedal switch failure, selecting
pedals allows the canopy to be raised or the bed to be raised or lowered.
4.4.3 General Information
The date of last time the temperature and line voltage calibration was performed and current time settings
appears across the bottom of the screen.
Last Cal: The date the System Calibration (Temperature and Line Voltage) was completed, the current date
and current time.
4.5 Setup
The setup screen allows you select basic device setup options. These options determine the items visible in
the user interface screens. Touch the pulldown and select the desired item.
Note:
If the values in the service mode setup screen are changed, the values are only implemented in the
main application when the patient history is cleared.
To access Setup features:
1.
In service mode, touch
Setup
in the left-navigation area.
The display area is populated with Setup options.
2.
Select desired settings. See the table below for detailed setting information.
A prompt appears as follows:
Applying Settings: Do not turn off power
.
FIGURE 4-3. Setup Screen
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...