Chapter 6: Alarms and System Failure Codes
System Failure Codes
Service Manual
6-13
6.2.2 System Failure 1
System failure 1 indicates:
No Timer2.
6.2.3 System Failure 2
System failure 2 indicates: Timer2 too fast.
Defective Servo Oxygen
cable
Disconnect RS485 bus cable from J6 on the control board. Power cycle the
device. If the error clears, replace the RS485 cable going to the Servo
Oxygen board.
Defective PF Battery
The battery is used only for power failure. Try powering up the device with
the battery disconnected. If the error clears, replace the battery.
Defective 14 pin ribbon
cable between control
board and SBC
Check that the 14 pin ribbon cable is properly seated at J6 on the control
board and at J23 on the SBC. (Have all three parts on hand and try them
one at a time. Try routing the cable around the outside of device first, then
try the SBC, and then try the control board. If the error is intermittent,
replace all three.)
Defective SBC
Replace SBC.
Defective control board
Replace control board.
Cause(s)
Action(s)
Cause(s)
Action(s)
The Timer2 circuit of the
microprocessor on the
control board is defective.
Replace the control board.
Cause(s)
Action(s)
Defective scale
Disconnect scale connector at probe panel. Power cycle the device. If the
error clears, replace the scale.
Defective scale harness
Disconnect scale harness from J7 on the control board. Power cycle the
device. If the error clears, replace the scale harness.
There is no line frequency
signal to feed timer0 and
timer1 of the
microprocessor on the
control board, or the
timer2 of the
microprocessor on the
control board, is defective.
Power cycle the device. If system failure 8 appears, refer to the instruction
in this chapter for system failure 8 (section
). If system failure 2
persists, replace the control board.
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...