Chapter 5: Troubleshooting
Status
5-22
Service Manual
5.4.4.4 Sensor Housing Temperature Sensor Temperature to Resistance Curve
5.5 Status
Access the Status screen from the main service menu. See “Accessing Service Mode” on page 4-1.
The Status screen displays status of all self tests the software runs continuously. These include: incubator
heater on (INCHTRON), warmer heater on (WRMHTRON), incubator/warmer heater off (I/WHTROFF),
humidifier heater on (RHHTR), remote monitoring data stream (RS-232LOOP), incubator fan on (FANON), and
incubator fan off (FANOFF). The current software revision of the options installed on this specific device
(Humidifier, Scale, SPO2, etc.) is shown.
If the RS-232 is not installed RS-232LOOP will display N/A. To test the circuit if the RS-232 is working, connect
the RS-232 to a working RS-232 compatible device or short pins 2 and 3 on the 9 pin connector.
Temperature
Resistance
15
°
C
15.720 kOhm
25
°
C
10.000 kOhm
30
°
C
8.056 kOhm
35
°
C
6.530 kOhm
40
°
C
5.326 kOhm
45
°
C
4.369 kOhm
50
°
C
3.604 kOhm
60
°
C
2.491 kOhm
FIGURE 5-2. Status Screen
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...