Chapter 5: Troubleshooting
Additional Troubleshooting Tips
Service Manual
5-11
The 5V and 12V supplies are generated by the Power Supply and are sent to the Relay Board for
distribution.The +5STBY is generated on the Relay Board from either the 12V (normal operation) or the 8.4V
NiMH PF Battery (power fail). These 3 voltages are distributed to the other boards via various cables.
The easiest place to measure a power supply voltage is at one of the spare 4 pin connectors on the
Expansion Slot Power Bus.
5.2.5 Troubleshooting Harness Cables 6600-0701-700 and 2075837-001
If DC power wire harness (6600-0701-700) or power supply DC output harness (2075837-001) is
disconnected or damaged, perform the following steps.
1.
Open the electronics enclosure on the back of the incubator and pull the control board out until you
expose the power connector (it is the 4-wire cable in the middle of the control board). See Figure 5-6.
FIGURE 5-6. Exposing the Power Connector
2.
Use a digital multimeter to measure the voltages between pin 1 and pins 2, 3 and 4, respectively.
Note:
Pin 1 is ground and is the top-most pin.
a.
Connect the negative (com) probe of multimeter to pin 1 and the positive probe to pin 2. The
voltage shall be more than 4.75V. If not, either harness 6600-0701-700 or harness 2075837-001 is
defective. Order new harnesses and replace them both.
b.
Connect the negative (com) probe of multimeter to pin 1 and the positive probe to pin 3. The
voltage shall be more than 11.5V. If not, either harness 6600-0701-700 or harness 2075837-001 is
defective. Order new harnesses and replace them both.
Board
Cable
Control Board
Expansion Slot Power Bus (6600-0701-700) from Relay Board
Servo Oxygen Board
Expansion Slot Power Bus (6600-0701-700) from Relay Board
RS-232 Board
RS-232 Cable (2074735-001) from Control Board
ICSBC
Rail Power Cable (6600-0706-700) from Expansion Slot Power Bus
Alarm Board
LINCAN Cable (2069666-001) from ICSBC
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...