4.1 Powering the Device
This section describes how to turn the device On and Off.
Isolation of the device from the supply mains can only be achieved by turning off the device using the mains
power switch on the rear of the device. The standby switch on the probe panel will not isolate the device
from the supply mains.
To turn the device On:
1.
Plug the power cord into a grounded outlet or, where available, a hospital grade outlet.
Note:
If the device has not been unplugged, verify that the device is plugged in and that the Mains power
switch is turned On.
2.
Turn on power at the Mains power switch on the back of the device.
3.
Turn on the power at the Standby switch on the probe panel.
When power is applied, the indicator light and Power Fail Indicator Light briefly turn on an then off. A
series of informational screens are displayed.
The device is ready for use when the Home screen appears.
To turn the device Off:
1.
Turn off the power at the Standby switch.
2.
If the device will be moved, turn off the Mains power switch at the back of the device, unplug the device,
and wrap the cord for safe transport.
3.
If the device will not be moved, no other action is needed for powering on or off.
4.2 Accessing Service Mode
By default, the device starts up in user mode. A series of user-hidden actions allows service personnel to
access Service Mode.
To access service mode:
1.
In user mode, touch
Setup
on the Home screen.
The Setup screen appears.
2.
Touch the data Baby Name field
Enter here
.
The keyboard appears.
FIGURE 4-1. User Setup Screen
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...