Chapter 3: Calibration and Maintenance
Electrical Safety Tests
Service Manual
3-9
For a successful calibration, the screen prompts
Calculating
for a few seconds.
14.
When the screen prompts
Exit Only
,
Restore Defaults
,
Save and Exit
, touch
Save and Exit
.
Note:
If
Save and Exit
is grayed out, the calibration was not successful. Repeat the calibration.
15.
Turn off the power to exit the service mode.
3.7 Electrical Safety Tests
Use an approved electrical safety analyzer to test the unit’s ground continuity and leakage current.
For reliable leakage current readings the ground check should be performed first.
Measurement of insulation resistance is not recommended. Long term effects of repeated exposure of high
voltage on the electrical insulation system of this medical device have not been fully evaluated.
FIGURE 3-8. Calculating
FIGURE 3-9. Completion
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...