Chapter 1: Functional Description
Humidifier
1-8
Service Manual
board, control flow to the patient compartment. Both valves are opened until the measured level gets close
to the desired set point then one valve is closed. One valve is then cycled on and off as needed to maintain
the desired oxygen levels in the patient compartment. The valve selected is alternated so both valves cycle
about the same number of times. There are 2 fuses between the Servo O2 board and the supply valves that
prevent high current from the board entering the valve housing should a short occur in the supply valves. A
safety valve that shuts off oxygen flow whenever the canopy is raised actuates mechanically by the
movement of the canopy support rail. When the canopy is down the valve is open (canopy up/valve closed).
The safety valve actuates independently of the solenoid type supply valves.
1.9 Humidifier
Refer to the following diagram. Water enters the insulating cylinder (1) through a small hole in bottom.
Steam rises along the heater cartridge protective insert (2) and is channeled between the heater mount (3)
and the ramp block (4) and out the steam port (5).
1.10 System Logs
All system failures are logged to the error log. The error logs are only accessible through the service
interface as described in the “Understanding the Service Interface” on page 4-1. The logs are stored to
persistent medium which preserves log data across power cycles and loss of power. When the log reaches
capacity, the oldest data is removed to provide space for newer data to be added.
3
2
1
5
4
FIGURE 1-2. Humidifier
Summary of Contents for Giraffe OmniBed Carestation CS1
Page 4: ...RH 2 Service Manual ...
Page 24: ...Service Language Disclaimer viii Service Manual ...
Page 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Page 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Page 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Page 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Page 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Page 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Page 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Page 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Page 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Page 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...